Table 1 Technical specific for JET 7 in comparison with the previous ACE 68, from Penumbra Reperfusion device family

From: Technical note and first results on JET 7 thromboaspiration device for T-ICA occlusions

Dimension

ACE™ 68

JET 7

Proximal OD

0.084 in Max

0.085 in Max

Proximal ID

0.068 in Min

0.072 in Min

Distal OD

0.084 in Max

0.085 in Max

Distal ID

0.068 in Min

0.072 in Min

Effective length

115, 120, 125, 127, 132 cm

Same

Distal flex length

30 cm

Same

Coating length

30 cm

Same

Indication

Penumbra Reperfusion Catheters and Separators As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for revascularization of patients with AIS secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Same