Table 1 qPCR assay validation parameters

From: Absence of lenadogene nolparvovec DNA in a brain tumor biopsy from a patient in the REVERSE clinical study, a case report

Matrix Calibration curve and regression

Herring sperm DNA, 400 ng per PCR well

Pass - Calibration standards ranging from 100 to 1 × 107 pAAV2/2-ND4 copies/well, fitted with a linear regression

Within-run precision and accuracy

Pass - CV% < 15.0, absolute RE% < 8.0

 

QC level

Nominal concentration (copies/well)

CV%

RE%

 

QC ULOQ

1.00E+ 07

11.1

−0.1

 

QC high

7.50E+ 06

8.0

−0.8

 

QC mid

1.05E+ 04

6.2

−3.5

 

QC low

3.00E+ 02

6.6

−7.2

 

QC LLOQ

1.00E+ 02

14.4

−2.0

Between run precision and accuracy

Pass - CV% < 9.0, absolute RE% < 6.0

 

QC level

Nominal concentration (copies/well)

CV%

RE%

 

QC ULOQ

1.00E+ 07

6.9

−3.1

 

QC high

7.50E+ 06

4.4

0.1

 

QC mid

1.05E+ 04

4.9

5.4

 

QC low

3.00E+ 02

8.2

4.2

 

QC LLOQ

1.00E+ 02

4.1

2.7

Dynamic range of the assay

100 to 1 × 107 copies/well corresponding to 250 to 2.5 × 107 copies/μg of DNA

Detection limit of the assay

6.25 copies/well corresponding to 15.625 copies/μg of DNA

Specificity of the assay

Pass - Tested with DNA extracted from blood samples from six humans

Matrix effect

Pass - Tested with DNA extracted from the six human blood samples spiked with 104 copies of pAAV2/2-ND4

  1. pAAV2/2-ND4 Plasmid containing the sequence for AAV2/2-ND4 DNA, CV% Percentage of Coefficient of variation, RE% Percentage of Relative Error, ULOQ Upper Limit of Quantification, LLOQ Lower Limit of Quantification.