Intracranial pressure responsiveness to positive end-expiratory pressure is influenced by chest wall elastance: a physiological study in patients with aneurysmal subarachnoid hemorrhage

Abstract

Background

Respiratory system elastance (ERS) is an important determinant of the responsiveness of intracranial pressure (ICP) to positive end-expiratory pressure (PEEP). However, lung elastance (EL) and chest wall elastance (ECW) were not differentiated in previous studies. We tested the hypothesis that patients with high ECW or a high ECW/ERS ratio have greater ICP responsiveness to PEEP.

Methods

An esophageal balloon catheter was placed to measure esophageal pressure. PEEP was increased from 5 to 15 cmH2O. Airway pressure and esophageal pressure were measured and EL, ECW and ERS were calculated at the two PEEP levels. Patients were classified into either an ICP responder group or a non-responder group based on whether the change of ICP after PEEP adjustment was greater than or less than the median of the overall study population.

Results

The magnitude of the increase in esophageal pressure (median [interquartile range]) at end-expiratory occlusion was significantly increased in the responder group compared with that in the non-responder group (4.1 [2.7–4.1] versus 2.7 [0.0–2.7] cmH2O, p = 0.033) after PEEP adjustment. ECW and the ECW/ERS ratio were significantly higher in ICP responders than in non-responders at both low PEEP (p = 0.021 and 0.017) and high PEEP (p = 0.011 and 0.025) levels. No significant differences in ERS and EL were noted between the two groups at both PEEP levels.

Conclusions

Patients with greater ICP responsiveness to increased PEEP exhibit higher ECW and a higher ECW/ERS ratio, suggesting the importance of ECW monitoring.

Peer Review reports

Background

Acute lung injury is prevalent in patients with acute brain injury [1,2,3,4]. Mechanical ventilation is needed in this population, and positive end-expiratory pressure (PEEP) is often used to improve oxygenation and prevent or recruit alveolar collapse [5,6,7]. However, there have long been concerns that the use of PEEP in brain-injured patients could reduce cerebral perfusion pressure due to both increased intracranial pressure (ICP) and decreased mean arterial pressure [8, 9]. Previous studies yielded inconsistent effects of PEEP on ICP [10] and diverse individual responsiveness [11,12,13], suggesting that the mechanism of action might be multifactorial. Several possible determinants for the influence of PEEP on ICP have been proposed, including baseline ICP [14], intracranial compliance [15, 16], respiratory mechanics [16, 17], dead space change and alveolar recruitment by PEEP [18].

Theoretically, PEEP may increase ICP by reducing venous return via elevating intrathoracic pressure [9]. Thus, ICP responsiveness to PEEP might largely depend on pressure transmission from the lung to the pleural cavity, which is determined by the distribution of lung elastance (EL) and chest wall elastance (ECW) in respiratory system elastance (ERS). Although clinical studies have suggested that an increased ERS might attenuate the effect of PEEP on ICP, EL and ECW were not differentiated in these studies [16, 17]. A given increased ERS might mainly contribute to the increase in EL due to pulmonary causes, such as acute respiratory distress syndrome (ARDS), or the increase in ECW due to the chest wall impairment, such as intra-abdominal hypertension or massive pleural effusion [19]. In mechanical ventilated patients, the reported ECW to ERS ratio (ECW/ERS ratio) varied from 0.2 to 0.8 [19]. To date, no study has been performed to determine the different effects of EL and ECW on ICP responsiveness to PEEP, which is worthy of investigation.

We hypothesized that the patients with high ECW or a high ECW/ERS ratio might exhibit more significant ICP responsiveness to PEEP. In the present study, patients with aneurysmal subarachnoid hemorrhage after clipping surgeries were enrolled, and two different levels of PEEP were applied. Changes in ICP after PEEP adjustment were monitored. Esophageal pressure was measured as a surrogate for pleural pressure, and the distribution of EL and ECW in ERS was determined. The aim was to investigate the possible influencing factors in ICP responsiveness to PEEP, especially for ECW and the ECW/ERS ratio.

Methods

Ethics and setting

This study was conducted in the Surgical Intensive Care Unit of Fujian Provincial Hospital, Fuzhou, China. The study protocol was approved by the Institutional Review Board of Fujian Provincial Hospital (K2015–023-01) on September 30, 2015, and the study was registered at ClinicalTrials.org (NCT02670733) on January 26, 2016 [20]. Because the study enrolled patients that were in a coma state, written informed consent was obtained from patient’s appropriate substitute decision maker designated to provide consent upon admission to the hospital.

Patients

All adult patients receiving aneurysm clipping after aneurysmal subarachnoid hemorrhage were consecutively screened daily. The inclusion criteria included 1) Glasgow Coma Scale (GCS) ≤ 8; 2) ventricular ICP monitor previously placed for ICP monitoring and cerebrospinal fluid drainage during the operation; and 3) need for mechanical ventilation with PEEP. The exclusion criteria were 1) age under 18 years; 2) after decompressive craniectomy; 3) ICP > 25 mmHg; 4) hemodynamic instability requiring greater than 10 μg/kg/min dopamine or more than 0.5 μg/kg/min norepinephrine; 5) history of esophageal surgery or chronic obstructive pulmonary disease; 6) evidence of active air leak from the lung or existing chest tube; and 7) expected to survive less than 24 h.

Given the lack of a widely accepted threshold to identify the responsiveness of ICP to increased PEEP levels, we classified each patient into two groups according to the median change of ICP after PEEP adjustment in the overall study population: above the median value was considered the “responder” group and below the median value was considered the “non-responder” group. This was an unblinded study, but the researchers were not aware of the patients’ allocation until all data were collected (the median can only be determined by then).

ICP monitoring and intracranial ventricular compliance measurements

ICP was measured using a ventricular ICP monitor (Codman, Johnson & Johnson, Raynham, MA, USA). Ventricular drainage was blocked during the procedure as long as tolerated by the patient (i.e. without an ICP > 25 mmHg). To measure the ventricular compliance, 2 mL of cerebrospinal fluid was drained, and immediate changes in ICP values were recorded. Ventricular compliance was calculated as the ratio of the cerebrospinal fluid drainage volume to the decrease in ICP after the drainage (mL/mm Hg) [15, 16].

Study procedure

During the study, the patient remained in a supine position with the head of the bed elevated to 30°. Esophageal pressure was measured by a SmartCath-G adult esophageal balloon catheter (7,003,300, CareFusion Co., Yorba Linda, CA, USA). An occlusion test was used to confirm the proper balloon position [21, 22]. Esophageal pressure and airway pressure were measured by two KT 100D-2 pressure transducers (KleisTEK di CosimoMicelli, Italy, range: +/− 100 cm H2O), and flow was measured with a Fleisch pneumotachograph (Vitalograph Inc., Lenexa, KS, USA). The pressure and flow signals were displayed continuously and saved (ICU-Lab 2.5 Software Package, ICU Lab, KleisTEK Engineering, Bari, Italy) on a laptop for further analysis. Details on esophageal pressure monitoring and data collection are presented in the Additional file 1.

After the establishment of esophageal pressure monitoring, the patient was sedated and paralyzed via the intravenous infusion of 5 mg of midazolam, 0.1 mg of fentanyl, and 50 mg of rocuronium. Mechanical ventilation was set as volume-controlled ventilation with a constant flow, an inspiratory to expiratory ratio of 1:2, and a tidal volume (VT) of 6 to 8 mL/kg of predicted body weight. The initial respiratory rate was set to maintain the PaCO2 at 35 to 45 mmHg. End-tidal carbon dioxide partial pressure (PETCO2) was also measured (CAPNOSTAT®, Maquet, Solna, Sweden). Pulse oxygen saturation was maintained above 92% by adjusting the FiO2.

Two PEEP levels were tested. The PEEP level was first set to 5 cm H2O. After a 30-min stabilization, PaO2, PaCO2 and PETCO2 were simultaneously measured, and the ratio of the alveolar dead space to the tidal volume (VDalv/VT) was calculated. ICP, mean blood pressure, cerebral perfusion pressure, central venous pressure and ventricular compliance were measured. End-expiratory and end-inspiratory occlusions were performed for 3 s each, and the airway pressure and the esophageal pressure during the last second of occlusion were recorded. Thereafter, the PEEP level was increased to 15 cm H2O and maintained for 30 min without changes in other ventilation settings. The same sequence of measurements was repeated, with the exception that the ventricular compliance measurement was not performed because cerebrospinal fluid drainage might influence ICP monitoring.

ERS, EL and ECW were calculated using a standard formula [19, 23]. The ECW/ERS ratio was documented. Detailed procedures, measurements and parameter derivations are presented in the Additional file 1.

The procedure was emergently terminated if ICP increased above 25 mmHg or cerebral perfusion pressure decreased below 50 mmHg.

Statistical analysis

The primary endpoint was the difference in ERS, EL, ECW and the ECW/ERS ratio in patients with different ICP responsiveness to PEEP. We did not calculate sample size because the differences in these elastance parameters between the two groups remain unknown. Instead, we chose 30, a widely accepted minimal sample size for a physiological study, as the sample size in the present study.

Categorical variables are presented as numbers and percentages and were analyzed by Fisher’s exact test. Continuous variables are presented as the median and inter-quartile range (IQR) and were compared using the Mann-Whitney U test or Wilcoxon matched-pair signed-rank test as appropriate. We used the Scheirer-Ray-Hare test to compare the parameters between the responder and the non-responder groups at low and high PEEP levels [24]. All tests of significance were at the 5% significance level and were two-sided. Analyses were performed with SPSS statistics software (V.20.0 IBM Corporation, New York, USA).

Results

From February to November 2016, 30 patients were studied. No emergent termination occurred during the procedures. Table 1 presents the main characteristics of the enrolled patients at the low (5 cm H2O) and high (15 cm H2O) PEEP levels. After the PEEP level was increased, ICP significantly increased (p <  0.001). The median (IQR) change in ICP was 2.5 (1.0–4.0) mm Hg. According to the predefined criterion, each of the 15 patients was allocated to the “responder” group or the “non-responder” group, which had respective ICP changes of 4.0 (3.0–5.0) and 1.0 (1.0–2.0) mm Hg. Baseline characteristics were comparable between the two groups except for responders with an older age and a lower body mass index (Table 2).

Table 1 Patients characteristics at low and high positive end-expiratory pressure
Table 2 Patients characteristics at study entry

Detailed data of respiratory mechanics, ICP and hemodynamic parameters and blood gas analysis are presented in the Additional file 2 (Table E1 to E6).

Esophageal pressure at end-expiratory occlusion increased significantly after PEEP adjustment (5.4 [4.1–8.2] versus 9.5 [6.8–12.2] cm H2O, p <  0.001), and the magnitude of the increase was significantly higher in the responder group than in the non-responder group (4.1 [2.7–4.1] versus 2.7 [0.0–2.7] cm H2O, p = 0.033, Fig. 1). At the low PEEP level, there were no significant differences in ERS (p = 0.136) and EL (p = 0.863) between two groups, but ECW and the ECW/ERS ratio were significantly higher in responders than in non-responders (p = 0.021 and 0.017) (Fig. 2a). As the PEEP level increased, no significant difference existed in the magnitude of change in ERS, EL, and ECW between the two groups (p = 0.468 to 0.787, Fig. 2b). Thus, significant differences in ECW and the ECW/ERS ratio between the two groups remained at the high PEEP level (p = 0.011 and 0.025, Fig. 2c).

Fig. 1
figure 1

End-expiratory esophageal pressure (PES-EE) measurements at different positive end-expiratory pressure (PEEP) levels. a The absolute values of PES-EE in the responder and non-responder groups. No significant difference was observed between the two groups. b The change of PES-EE during the adjustment of PEEP. The change of PES-EE in the responders was significantly increased compared with that in non-responders, suggesting that more pressure was transmitted to the pleural cavity

Fig. 2
figure 2

Elastance of the lung, the chest wall and the respiratory system at different positive end-expiratory pressure (PEEP) levels. a Elastance measured at a PEEP level of 5 cm H2O. Chest wall elastance (ECW) was significantly increased in responders compare with that in non-responders. There were no significant differences in respiratory system elastance (ERS) and lung elastance (EL) between the two groups. b The change of EL, ECW and ERS during the adjustment of PEEP. No significant difference was observed between the two groups. c EL, ECW and ERS measured at a PEEP level of 15 cm H2O. ECW remained significantly increased in responders compared with that in non-responders. There were no significant differences in ERS and EL between the two groups

No significant difference was found in baseline ICP (3.0 [2.0–6.0] versus 4.0 [4.0–10.0] mm Hg, p = 0.163) and intracranial ventricular compliance (1.0 [0.7–1.0] versus 1.0 [0.7–1.0] mL/mm Hg, p = 0.723) between the ICP responder and non-responder groups. Although central venous pressure increased significantly after the PEEP level was increased in all enrolled patients (Table 1, p <  0.001), no significant difference was observed in the change of central venous pressure between ICP responders and non-responders (5.0 [4.0–5.0] versus 3.0 [1.0–6.0] mm Hg, p = 0.077, Fig. 3). Similarly, significant decreases in mean blood pressure were observed after PEEP increase in all enrolled patients (81.0 [75.6, 92.1] versus 76.2 [67.4, 90.0], p = 0.002), but no difference in the change of mean blood pressure between groups (Table E3 and E4). Cerebral perfusion also decreased after PEEP was increased, the responder had a significantly greater decrease in cerebral perfusion pressure than the non-responder (− 13.0 [− 18.3, − 6.7] versus − 6.7 [− 11.3, 1.0], p = 0.011; Table E3 and E4).

Fig. 3
figure 3

Central venous pressure (CVP) at different positive end-expiratory pressure (PEEP) levels. a The absolute values of CVP in the responder and non-responder groups. b The change of CVP during the adjustment of PEEP

There was no significant difference in PaO2, PaO2/FiO2 and PaCO2 after PEEP adjustment (p = 0.052 to 0.061, Table 1). PETCO2 significantly decreased (29.5 [27.8–31.2] versus 29.0 [25.8–30.3] mm Hg, p = 0.003); thus, VDalv/VT significantly increased (13.2 [8.5–19.9] versus 18.8 [11.9–27.7] %, p <  0.001). No significant differences were observed in PETCO2 and VDalv/VT between ICP responders and non-responders at either the low or high PEEP levels.

Discussion

The main finding of our study was that patients with a greater ICP responsiveness to increased PEEP exhibit a higher ECW and ECW/ERS ratio. This finding was consistent with our hypothesis.

Regarding the influence of PEEP on ICP, clinical studies have reported conflicting results, with ICP increasing [14,15,16,17,18], not markedly changing [25,26,27] or even decreasing [28] after the application of PEEP. In patients with increased ICP, the effect of PEEP on ICP becomes evident whenever the elevation of the intrathoracic pressure induced by applied PEEP exceeds the ICP [14]. This mechanism was referred to as the Starling resistor model, which represented the role of damaged venous outflow in the response of ICP to PEEP [29]. In a study conducted by McGuire et al., 15 mmHg was used as the upper limit of the normal baseline value of ICP at zero PEEP [14]. ICP increased at PEEP levels of 10 and 15 cm H2O in patients with normal baseline ICP but did not significantly change at PEEP levels of 5 to 15 cm H2O in patients with elevated baseline ICP. In our group of patients, all ICP values were less than 15 mmHg at 5 cm H2O PEEP (4.0 [2.0–10.0] mm Hg) and increased after the PEEP level was adjusted to 15 cm H2O (7.0 [5.8–11.0] mm Hg). The difference in ICP at the two PEEP levels (2.5 [1.0–4.0] mm Hg) in our patients was comparable to those previously reported [13, 14].

The PEEP-induced increase of pleural pressure may reduce cerebral venous drainage and eventually increase ICP [9]. The pleural pressure serves as an intermediate link from the lung to the cranium. In the present study, we measured esophageal pressure as a surrogate of pleural pressure and found that esophageal pressure increased more significantly in ICP responders than in non-responders (Fig. 1). This finding suggested a role of pressure transmitted to the pleural cavity in ICP responsiveness. In addition, central venous pressure increased significantly at the high PEEP level (Table 1), indicating the impairment of venous return by PEEP, which eventually resulted in reduced mean blood pressure and cerebral perfusion pressure. ICP responders had a higher tendency to have elevated central venous pressure (Fig. 3); however, statistical significance was not achieved, potentially due to the limited sample size.

The elastic properties of the respiratory system may be a key factor in determining the effect of PEEP on pleural pressure and subsequently ICP [9]. Pressure transmission from the lung to the pleural cavity depends on the distribution of EL and ECW in ERS. The higher the proportion of ECW is, the greater the impact of PEEP on pleural pressure [19]. Therefore, ECW and the ECW/ERS ratio may be a more important determinant of ICP responsiveness to PEEP. In previous clinical studies, only ERS was used to explore the influence of respiratory mechanics in the PEEP and ICP relationship [16, 17]. After distinguishing EL and ECW from ERS, we found that the effect of PEEP on ICP was more profound in patients with higher ECW and a higher ECW/ERS ratio (Fig. 2). Although patients with brain injury exhibited higher ECW (without detailed values) in previous studies [30], ECW and the ECW/ERS ratio measured in our group of patients were relatively “normal” (Table 1). Our results were comparable to those reported in patients under general anesthesia [31] but higher than those reported in ARDS patients [32]. However, even a slightly higher ECW and ECW/ERS ratio might contribute to ICP responsiveness (Fig. 2). Our data suggested the importance of separate EL and ECW monitoring during the clinical application of high PEEP in brain-injured patients.

There are two consequences after the application of PEEP: recruitment of collapsed alveoli and/or overdistention of normal alveoli. In patients with severe brain injury and acute lung injury or ARDS, Mascia et al. defined two groups based on recruitment volume: ≥ 110 mL as recruiters and < 110 mL as non-recruiters [18]. As the PEEP level increased, ERS and PaCO2 significantly increased in non-recruiters, indicating hyperinflation. On the contrary, ERS significantly decreased, and PaCO2 remained unchanged in recruiters. ICP increased only in non-recruiters. In their patient population, the PaO2/FiO2 ratio was 95–141 mmHg, and ERS was 24–25 cm H2O/L at 5 cm H2O PEEP [18]. In our group of patients, only six ARDS cases (20%) were enrolled, and the overall severity of lung injury was mild, as represented by the median PaO2/FiO2 ratio of 264 and an ERS of 15.3 cm H2O/L at the low PEEP level. Alveolar collapse might also be minor; therefore, increases in the PEEP level produced increases in ERS (15.3 to 20.1 cm H2O/L) and VDalv/VT (13.2 to 18.8%). Despite lacking statistical significance, PaCO2 and VDalv/VT were high in ICP responders. Therefore, PEEP-induced hyperinflation might also contribute to ICP responsiveness in our patients.

The effect of PEEP on ICP might be affected by ventricular compliance [15, 16]. Apuzzo et al. reported that a significant increase in ICP was only observed in patients who manifested increased cerebral elastance when PEEP was applied [15]. Similarly, Burchiel et al. reported that PEEP increased ICP in patients with decreased intracranial compliance and that decreased lung compliance may buffer this effect [16]. In the present study, given that patients with normal ICP were enrolled and that cerebrospinal fluid was continuously drained before the start of the study, the ventricular compliance was “normal” in all patients. Additionally, for ethical reasons, we cannot measure the ventricular compliance by injecting normal saline into the ventricle instead of draining the cerebrospinal fluid. Because intracranial hypertension and ventricular compliance impairment are prevalent in patients with severe brain injury, further animal experiments should be performed to validate the role of ventricular compliance in the ICP and PEEP relationship.

Increase in PEEP decreases systematic venous return and potentially lowers the cardiac output, this will lower blood pressure and cerebral perfusion pressure. If the cerebral perfusion pressure break through the lower limit of cerebral autoregulation, cerebral perfusion will not be maintained and thus decrease cerebral blood flow, cerebral blood volume and ICP. In this case, the decrease of cerebral perfusion (and thereby decrease in ICP) will somehow contaminate the effect of PEEP on the venous side. In our study, blood pressure and cerebral perfusion pressure were maintained in a normal range, even in a high PEEP level. Therefore, although without cerebral blood flow being measured, we assumed that the cerebral perfusion was comparable between the two groups.

The responders were significant older in this study. It has been reported that the elders have higher ECW due to structural changes to the thoracic cage [33, 34], which is the result of age-related osteoporosis and calcification of the rib cage that reduce the ability of the thoracic cage to expand. Our data also showed higher ECW in the elders, probably due to the same reason. The responders also had significantly lower BMI. Although obesity people have higher ERS, it can be the result of either increased EL or increased ECW, or even both [35]. This makes the ECW unpredictable simply base on BMI. Moreover, our patients had relatively normal BMI, the small difference of BMI between the two group (although statistically significant) is unlikely the determinant of the difference of ECW.

The main strength of the present study was that the contribution of the chest wall and the lung in ICP responsiveness to PEEP was distinguished. The adverse effect of PEEP on ICP may be amplified in patients with higher ECW and a higher ECW/ERS ratio. Our data suggest the potential necessity of respiratory mechanics monitoring when applying PEEP in brain-injured patients, especially in patients with a risk of chest wall impairment.

There were several limitations of the study. First, we used the median of the increased value rather than other physiological parameters to define the ICP responders and non-responders. Given that no widely accepted standard is available to discriminate whether the patient’s ICP is responsive to increased PEEP, the division of patients into two groups is reasonable and enables us to compare potential determinants between the two groups. Second, we only included postoperative subarachnoid hemorrhage patients with relatively normal ICP and respiratory mechanics. Therefore, our results might not be directly applied to other populations. Third, although esophageal pressure has been employed as a surrogate of pleural pressure, the use of the absolute value of esophageal pressure remains questionable [19]. However, in the calculation of ECW, we used the change in esophageal pressure between end-inspiration and end-expiration occlusion, which has been shown to reliably reflect the change in pleural pressure [23, 36, 37].

Conclusions

Patients with greater ICP responsiveness to increased PEEP had higher ECW and a higher ECW/ERS ratio. Potential factors contributing to the increase in ECW and the ECW/ERS ratio should be assessed and eliminated if possible to avoid the adverse effects of high PEEP levels on the brain.

Abbreviations

ARDS:

Acute respiratory distress syndrome

CPP:

Cerebral perfusion pressure

CVP:

Central venous pressure

ECW :

Chest wall elastance

EL :

Lung elastance

ERS :

Respiratory system elastance

ICP:

Intracranial pressure

MAP:

Mean arterial pressure

PAW :

Airway pressure

PEEP:

Positive end-expiratory pressure

PES :

Esophageal pressure

PETCO2 :

End-tidal carbon dioxide partial pressure

VDalv/V T :

Alveolar dead space to the tidal volume ratio

VT :

Tidal volume

References

  1. Hoesch RE, Lin E, Young M, Gottesman RF, Altaweel L, Nyquist PA, et al. Acute lung injury in critical neurological illness. Crit Care Med. 2012;40:587–93.

    Article  PubMed  Google Scholar 

  2. Kahn JM, Caldwell EC, Deem S, Newell DW, Heckbert SR, Rubenfeld GD. Acute lung injury in patients with subarachnoid hemorrhage: incidence, risk factors, and outcome. Crit Care Med. 2006;34:196–202.

    Article  PubMed  Google Scholar 

  3. Aisiku IP, Yamal JM, Doshi P, Rubin ML, Benoit JS, Hannay J, et al. The incidence of ARDS and associated mortality in severe TBI using the berlin definition. J Trauma Acute Care Surg. 2016;80:308–12.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Mascia L, Sakr Y, Pasero D, Payen D, Reinhart K, Vincent JL. Extracranial complications in patients with acute brain injury: a post-hoc analysis of the SOAP study. Intensive Care Med. 2008;34:720–7.

    Article  PubMed  Google Scholar 

  5. Pelosi P, Ferguson ND, Frutos-Vivar F, Anzueto A, Putensen C, Raymondos K, et al. Management and outcome of mechanically ventilated neurologic patients. Crit Care Med. 2011;39:1482–92.

    Article  PubMed  Google Scholar 

  6. Tejerina E, Pelosi P, Muriel A, Penuelas O, Sutherasan Y, Frutos-Vivar F, et al. Association between ventilatory settings and development of acute respiratory distress syndrome in mechanically ventilated patients due to brain injury. J Crit Care. 2017;38:341–5.

  7. Luo XY, Hu YH, Cao XY, Kang Y, Liu LP, Wang SH, et al. Lung-protective ventilation in patients with brain injury: a multicenter cross-sectional study and questionnaire survey in China. Chin Med J. 2016;129:1643–51.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Mascia L. Ventilatory setting in severe brain injured patients: does it really matter? Intensive Care Med. 2006;32:1925–7.

    Article  PubMed  Google Scholar 

  9. Koutsoukou A, Katsiari M, Orfanos SE, Kotanidou A, Daganou M, Kyriakopoulou M, et al. Respiratory mechanics in brain injury: a review. World J Crit Care Med. 2016;5:65–73.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Borsellino B, Schultz MJ, Gama de Abreu M, Robba C, Bilotta F. Mechanical ventilation in neurocritical care patients: a systematic literature review. Expert Rev Respir Med. 2016;10:1123–32.

    Article  PubMed  CAS  Google Scholar 

  11. Shapiro HM, Marshall LF. Intracranial pressure responses to PEEP in head-injured patients. J Trauma. 1978;18:254–6.

    Article  PubMed  CAS  Google Scholar 

  12. Georgiadis D, Schwarz S, Baumgartner RW, Veltkamp R, Schwab S. Influence of positive end-expiratory pressure on intracranial pressure and cerebral perfusion pressure in patients with acute stroke. Stroke. 2001;32:2088–92.

    Article  PubMed  CAS  Google Scholar 

  13. Zhang XY, Yang ZJ, Wang QX, Fan HR. Impact of positive end-expiratory pressure on cerebral injury patients with hypoxemia. Am J Emerg Med. 2011;29:699–703.

    Article  PubMed  Google Scholar 

  14. McGuire G, Crossley D, Richards J, Wong D. Effects of varying levels of positive end-expiratory pressure on intracranial pressure and cerebral perfusion pressure. Crit Care Med. 1997;25:1059–62.

    Article  PubMed  CAS  Google Scholar 

  15. Apuzzo JL, Wiess MH, Petersons V, Small RB, Kurze T, Heiden JS. Effect of positive end expiratory pressure ventilation on intracranial pressure in man. J Neurosurg. 1977;46:227–32.

    Article  PubMed  CAS  Google Scholar 

  16. Burchiel KJ, Steege TD, Wyler AR. Intracranial pressure changes in brain-injured patients requiring positive end-expiratory pressure ventilation. Neurosurgery. 1981;8:443–9.

    Article  PubMed  CAS  Google Scholar 

  17. Caricato A, Conti G, Della Corte F, Mancino A, Santilli F, Sandroni C, et al. Effects of PEEP on the intracranial system of patients with head injury and subarachnoid hemorrhage: the role of respiratory system compliance. J Trauma. 2005;58:571–6.

    Article  PubMed  Google Scholar 

  18. Mascia L, Grasso S, Fiore T, Bruno F, Berardino M, Ducati A. Cerebro-pulmonary interactions during the application of low levels of positive end-expiratory pressure. Intensive Care Med. 2005;31:373–9.

    Article  PubMed  Google Scholar 

  19. Gattinoni L, Chiumello D, Carlesso E, Valenza F. Bench-to-bedside review: chest wall elastance in acute lung injury/acute respiratory distress syndrome patients. Crit Care. 2004;8:350–5.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Chen H, Xu M, Yang YL, Chen K, Xu JQ, Zhang YR, et al. Effects of increased positive end-expiratory pressure on intracranial pressure in acute respiratory distress syndrome: a protocol of a prospective physiological study. BMJ Open. 2016;6:e012477.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Baydur A, Behrakis PK, Zin WA, Jaeger M, Milic-Emili J. A simple method for assessing the validity of the esophageal balloon technique. Am Rev Respir Dis. 1982;126:788–91.

    PubMed  CAS  Google Scholar 

  22. Chiumello D, Consonni D, Coppola S, Froio S, Crimella F, Colombo A. The occlusion tests and end-expiratory esophageal pressure: measurements and comparison in controlled and assisted ventilation. Ann Intensive Care. 2016;6:13.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Mauri T, Yoshida T, Bellani G, Goligher EC, Carteaux G, Rittayamai N, et al. Esophageal and transpulmonary pressure in the clinical setting: meaning, usefulness and perspectives. Intensive Care Med. 2016;42:1360–73.

    Article  PubMed  Google Scholar 

  24. Scheirer CJ, Ray WS, Hare N. The analysis of ranked data derived from completely randomized factorial designs. Biometrics. 1976;32:429–34.

    Article  PubMed  CAS  Google Scholar 

  25. Frost EA. Effects of positive end-expiratory pressure on intracranial pressure and compliance in brain-injured patients. J Neurosurg. 1977;47:195–200.

    Article  PubMed  CAS  Google Scholar 

  26. Cooper KR, Boswell PA, Choi SC. Safe use of PEEP in patients with severe head injury. J Neurosurg. 1985;63:552–5.

    Article  PubMed  CAS  Google Scholar 

  27. Koutsoukou A, Perraki H, Raftopoulou A, Koulouris N, Sotiropoulou C, Kotanidou A, et al. Respiratory mechanics in brain-damaged patients. Intensive Care Med. 2006;32:1947–54.

    Article  PubMed  Google Scholar 

  28. Huynh T, Messer M, Sing RF, Miles W, Jacobs DG, Thomason MH. Positive end-expiratory pressure alters intracranial and cerebral perfusion pressure in severe traumatic brain injury. J Trauma. 2002;53:488–92. discussion 92-3

    Article  PubMed  Google Scholar 

  29. Luce JM, Huseby JS, Kirk W, Butler J. A Starling resistor regulates cerebral venous outflow in dogs. J Appl Physiol Respir Environ Exerc Physiol. 1982;53:1496–503.

    PubMed  CAS  Google Scholar 

  30. Gamberoni C, Colombo G, Aspesi M, Mascheroni C, Severgnini P, Minora G, et al. Respiratory mechanics in brain injured patients. Minerva Anestesiol. 2002;68:291–6.

    PubMed  CAS  Google Scholar 

  31. Cinnella G, Grasso S, Spadaro S, Rauseo M, Mirabella L, Salatto P, et al. Effects of recruitment maneuver and positive end-expiratory pressure on respiratory mechanics and transpulmonary pressure during laparoscopic surgery. Anesthesiology. 2013;118:114–22.

    Article  PubMed  CAS  Google Scholar 

  32. Chiumello D, Colombo A, Algieri I, Mietto C, Carlesso E, Crimella F, et al. Effect of body mass index in acute respiratory distress syndrome. Br J Anaesth. 2016;116:113–21.

    Article  PubMed  CAS  Google Scholar 

  33. Sharma G, Goodwin J. Effect of aging on respiratory system physiology and immunology. Clin Interv Aging. 2006;1:253–60.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  34. Mittman C, Edelman N, Norris a SN. Relationship between chest wall and pulmonary compliance with age. J Appl Physiol. 1965;20:1211–6.

    Article  Google Scholar 

  35. Littleton SW. Impact of obesity on respiratory function. Respirology. 2012;17:43–9.

    Article  PubMed  Google Scholar 

  36. Chiumello D, Carlesso E, Cadringher P, Caironi P, Valenza F, Polli F, et al. Lung stress and strain during mechanical ventilation for acute respiratory distress syndrome. Am J Respir Crit Care Med. 2008;178:346–55.

    Article  PubMed  Google Scholar 

  37. Chiumello D, Cressoni M, Colombo A, Babini G, Brioni M, Crimella F, et al. The assessment of transpulmonary pressure in mechanically ventilated ARDS patients. Intensive Care Med. 2014;40:1670–8.

    Article  PubMed  Google Scholar 

Download references

Funding

The study was funded by the Beijing Municipal Administration of Hospital (grant number ZYLX201502). The sponsors had no role in the design of the study and collection, analysis, and interpretation of data and in writing the manuscript.

Availability of data and materials

The datasets used and/or analyzed during the present study are available from the corresponding author on reasonable request.

Author information

Authors and Affiliations

Authors

Contributions

HC, JXZ and RGY participated in the design of the study and drafted the manuscript. HC, KC, JQX and YRZ participated in the study and data collection. All authors participated in review and revision of the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Jian-Xin Zhou.

Ethics declarations

Ethics approval and consent to participate

The study protocol was approved by the Institutional Review Board of Fujian Provincial Hospital (K2015–023-01) on September 30, 2015. Because the study enrolled patients that were in a coma state, written informed consent was obtained from patient’s appropriate substitute decision maker designated to provide consent upon admission to the hospital.

Consent for publication

Not applicable

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Additional files

Additional file 1:

Detailed methods of the study procedures, measurements and parameter derivations. (PDF 186 kb)

Additional file 2:

Detailed results of respiratory mechanics, intracranial pressure, hemodynamics parameters and blood gas. Values of respiratory mechanics (Table E1), intracranial pressure and hemodynamics parameters (Table E3) and blood gas (Table E5) at low and high positive end-expiratory pressure (PEEP) levels were provided. Changes in respiratory mechanics (Table E2), intracranial pressure and hemodynamics parameters (Table E4) and blood gas (Table E6) from low to high PEEP were also provided. (PDF 177 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Chen, H., Chen, K., Xu, JQ. et al. Intracranial pressure responsiveness to positive end-expiratory pressure is influenced by chest wall elastance: a physiological study in patients with aneurysmal subarachnoid hemorrhage. BMC Neurol 18, 124 (2018). https://doi.org/10.1186/s12883-018-1132-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12883-018-1132-2

Keywords